Fuel Cell Technology
Digital Alcohol Detector
Part Names
This device is a breath alcohol tester designed to measure the presence of alcohol in the body. Please make sure to read this manual thoroughly before use.
DO NOT DRINK AND DRIVE
Display
Operation
1. Insert a mouthpiece into the mouthpiece inlet.
2. Push the ON/OFF power switch that is located in the right side under the display window.
- The unit will show "Full display" firstly and "on" along with a welcome sound. And then it starts countdown which will be taken about 15 ~ 35 seconds. This warm-up process prepares the sensor and circuit for testing.
3. WHen you hear a beep sound & "bLo" display is on, please blow steadily into the mouthpiece until you would hear the "Beep" sound again. Stop blowing.
- In this stage, if you don't blow within 20 seconds, it automatically turns off. Try to power on again.
4. After the circling line is on the display for 3 ~ 22 seconds, the test result (%BAC) will be displayed for 20 seconds along with beep sound.
- If the test result is over 0.050%BAC (default setting) respectively, "WARN" signal (Red LED) will be displayed along with "Alarm" sound.
- The result will be displayed for 20 seconds (Result display along with "READY" or "WARN" LED on for the first 15 seconds and LED off for another 5 seconds). If you push the power switch when the Green or Red LED is off, the unit goes back to countdown stage for another test.
5. Finally, the unit will show "OFF" and turn off automatically along with a sound.
6. For the successive testing, try again from No.2 to No.4
7. Space tests at least 2 minutes apart.
Error(Warning) Notification
Error
Message |
Cause & Action
|
Error
Message |
Cause & Action
|
![]() |
If the result is higher than 0.050% BAC, "WARN" signal is displayed along with "ALARM" sound. |
![]() |
The battery voltage is too low to perform a test. The device will turn off automatically after showing this signal. Replace 9V Alkaline Battery. |
![]() |
If the result is above 0.400% BAC, "Hot" signal is displayed along with alarm sound. |
![]() |
The user had a short exhale time during the test, which likely means the user did not exhale steadily and continuously for at least 3~5 seconds. Push the power button to perform another test. |
![]() |
When the test time is over 500, the unit shows "CAL". Even if the unit still works, it is required the recalibration. Consult to the seller or local distributor for the service. |
![]() |
When the ambient temperature is out of working temperature range (5°C ~ 40°C) the unit shows "OUT" and then turn off. Keep and use the unit within the recommended temperature range. |
Battery Replacement
Apply a small amount of pressure on the battery cover located on the back of the unit and push down and away from the unit to release the cover. See the following diagram.
** N.B. : Make sure to insert the battery with the proper position of positive(+) and negative(-) as indicated.
Precautions
1. After drinking, smoking and eating, it is recommended to test after 20 min.
- It is because it takes 20 min. approximately for alcohol to be absorbed into blood from the digestive organs, and residual alcohol remaining in the mouth takes this long to dissipate.
2. Avoid testing under strong wind or in the closed room with contaminated air.
3. When the "BAT" signal is on the display, it is advisable to replace 9V alkaline battery.
4. Do not blow cigarette smoke, food or liquid into the instrument because this may damage the sensor.
5. Do not use this unit as a tool to determine whether you or anyone else should operate any motor vehicle or machinery.
6. Keep it under 0. 40°C (32. 104°F). And this device is designed to be used in a temperature range of 5. 40°C (41. 104°F). Keep and use the unit within the recommended temperature range.
7. Correlation between breath alcohol content and blood alcohol concentration depends on many variables, including environmental factors (such as air quality, wind, humidity, temperature, etc.) and health conditions of the test subject.
8. In normal use, CA20FS requires re-calibration every 6 months or after 500 tests to obtain accurate results. Discuss with local distributors for re-calibration.
9. You cannot use the results of this unit in court of law. The driver holds full responsibility for driving under the influence of alcohol.
10. Do not use this device in conjunction with an ignition interlock device or while sitting in a motor vehicle .
11. An individual's testing result can continue to increase or decrease over time, depending on many factors. Obtaining a testing result reading at one time does not mean the testing result reading will be the same even a few minutes later.
12. Do not let yourself or anyone else drink and drive. Always use a designated driver when you or anyone else drinks alcohol.
Specification
Sensor
|
Smart Fuel Cell Sensor
|
Accuracy
|
±0.005% BAC at 0.050% BAC
|
Detection Range
|
0.000 ~ 0.400% BAC
|
Warm-up Time
|
15~35 seconds
|
Response Time
|
3~22 seconds
|
Dimension
|
125mm x 56mm x 24mm
|
Weight
|
87g, (131g Battery included)
|
Power Supply
|
9V alkaline battery
|
Working Temp.
|
5~40°C
|
Storage Temp.
|
0~40°C
|
Calibration
|
Every 6 months or after 500 tests
|
Warranty
|
1 Year Limited Warranty
|
Registration Date | 2014/04/06(Year/Month/Date) |
---|---|
Buyer / Seller in EC21 | Seller |
Business Type | Manufacturer |
Year established | 2011 |
Employees total | 6 - 10 |
Annual revenue | Less than USD 100,000 |
Company | C&L Korea |
---|---|
Address | 19,23Gil, World cup Ro,Mapo-GuSeoul121-820Korea |
Phone | 82 - 2 - 5620107 |
Fax | 82 - 2 - 3350107 |
Contact | Charles Lee / CEO |